Get Permission Sundaram, Dhakshinamoorthy, and Swaminathan: Observational study on the accuracy and speed of ultrasound guided transversus abdominis plane block with two different monitor positions


Introduction

Ultrasound guided regional anaesthesia (UGRA) has gained popularity during the last decade. Its main advantage is the ability to identify anatomical structures in real time; needle trajectory and injectate around the nerve structures.1 Meta analysis have shown that, the overall success rate of UGRA is higher when compared to other methods.2 UGRA procedures are fast, quick in onset, provide prolonged action, with low risk of vascular punctures and local anaesthetic toxicity.3 Some of the prerequisites of successful UGRA blockade include knowledge of human anatomy and principles related to USG blocks such as good hand skills, hand eye coordination and proper position of monitor and needle placements. Educating and training anaesthesiologists to the highest standards is of paramount importance to deliver high such quality care.

This prompted us to conduct an observational study on the accuracy and speed of ultrasound guided transversus abdominis plane block (TAP) with two different monitor position for hernia surgeries. Inguinal hernia repair is one of the most commonly performed surgical procedures worldwide.4 With the advent of UGRA techniques, the TAP block has proven to be effective in reducing acute post operative pain.5 TAP blocks have been described as an effective component of multimodal post operative analgesia for various abdominal procedures like hysterectomy, appendicectomy, colorectal surgeries, TURP and inguinal hernia. Controlling acute post operative pain reduces the incidence of chronic pain development.6

Materials and Methods

After approval of hospital ethics committee and after obtaining written informed consent from anaesthetists and patients, a prospective observational study on the accuracy and speed of USG TAP block with two different monitor positions is conducted. In first group monitor was kept in line of sight and in second group monitor was kept perpendicular to the participant. Twelve anaesthetists performed one procedure in each group. Anaesthetists who had performed more than 20 procedures were included in the study. History of back and neck pain was elicited and any discomfort during the procedure was recorded. Twenty four male patients of age 18-60 years of ASA I & II posted for inguinal hernia repair under spinal anaesthesia were included in the study. Patient refusal, BM1>30 kg/m2, bilateral hernias were excluded from the study. Recruitment was dependent on the patient and anaesthesiologists consent and availability of investigator who was responsible for data collection.

All the procedures were done under spinal anaesthesia under standard sterile protocol. On the completion of surgical procedure USG TAP block was performed with standard sterile precautions. Heart rate, oxygen saturation, non invasive blood pressure monitoring, electrocardiograph were monitored during the procedure. After the identification of iliac crest and sub costal margin, linear probe of frequency 6-14 MHz (SonoSite SII) was placed in midaxillary line on a transverse plane. After visualization of abdominal layers, 23G 50 mm insulated nerve block needle was inserted 1 cm medial to the probe and was advanced using in plane technique. Observation period started from probe positioning to injection of 20 ml of 0.25% bupivacaine between internal oblique and transversus abdominis muscles. Quality of image was scored from 1 to 4 as described by Oliveira7 as shown in Table 1. Statistical analysis was performed using statistical package for social sciences (SPSS).

Table 1

Image of accuracy score by Oliveira

1 Ideal – The needle was in full visualization during its progression.
2 Good – Needle’s tip was visible during the entire progression but the shaft was only partially visible.
3 Satisfactory – Only the needle’s tip was visible.
4 Poor – Only tissue distortion could be observed during the needle progression.

Results

Twelve anaesthetists were recruited and performed one procedure in each group. Time taken by anaesthetists to perform the block in group I was 69.08 ± 8.19 Sec. whereas in Group II was 80.6 ± 4.84 Sec. as shown in Table 2. The difference was statistically significant between Group I and II (P< 0.00) as shown in Figure 1.

Table 2

Comparison of duration of procedure and accuracy ofneedle placement between group I and group II

Group I Group II P value
Duration of procedure (Sec.) 69.08 ± 8.19* 80.6 ± 4.84* < 0.001(s)
Total image accuracy score 15 24 <0.01 (s)

[i] * Values are expressed as mean ± SD; S – Significant

Figure 1

Comparison of duration of procedure between group I and group II

https://typeset-prod-media-server.s3.amazonaws.com/article_uploads/703d8b84-8d4d-4d59-bece-712c67381cf7/image/dac0940b-5631-4d7c-a69f-f31e13352f53-uimage.png

Image accuracy score was ideal nine times in Group I two times in Group II; score was good 3 times in Group I and 8 times in Group II; Satisfactory 2 times in Group II as shown in Table 3. Total image accuracy score in Group I was 15 whereas in Group II it was 24. Chi-square test was applied. Critical value of chi-square distribution with df 2 is 9.210 which was more than the calculated value 8.727 (<0.001 statistically significant). Out of 12 anaesthetists, 4 had history of neck pain previously and one was on the treatment for the same. When performed in Group I, they had no discomfort, but when performed in Group II, 2 anaesthetists with positive history had discomfort during the procedure. All the blocks were successful.

Table 3

Image accuracy score

Group Ideal Good Satisfactory Poor
Group I 9 3
Group II 2 8 2

Discussion

Our study indicated that performance of USG TAP block was comfortable and more accurate and procedure time was shorter, if the monitor was kept in front. This was supported by Chapman,8 who stated that the anatomical area and screen should be in same line of view during the procedure. He also stressed the importance of keeping the screen at the eye level of operator.

Langford9 et al. who studied the performance of 31 anaesthetists on phantom with two monitor positions, found that the accuracy of needle placement was more when monitor was kept in front. He also stressed the importance of accuracy of needle placement not only to increase the success of blockade but also to reduce accidental damage to other nearby structures.

Speer10 in his study described two ways of placing transducer while performing task on pork phantom model, where the monitor was kept in front during the procedures. He found that there was reduction in time to perform the task and the quality of needle imaging was better when the transducer was placed along the visual axis (ALVA) compared to transducer kept across the visual axis (ACVA). He also suggested that this ergonomic aspect will be appropriate for novices learning USG procedures.

This is in accordance with our study, where duration of procedure was short when the monitor was in line with visual axis. But it was contradicted by Langford9 who stated that for anaesthetists with limited USG experience speed was not improved by aligning the monitor in line of sight of operator.

Wilson11 compared ALVA and ACVA techniques in medical students on phantom, found that ALVA techniques minimizes the needle passes, decrease the time required for task completion. He also stressed the ALVA techniques enhance the patient safety and outcome when performed by trainees. And it improved the ergonomic performance of the task which was also preferred by trainees. In our study 2 of the 4 anaesthetists with positive neck pain history had mild discomfort during the procedure. They preferred, the monitor in front procedures. Maintaining a bad posture is the source of muscle fatigue and it reduces the performance of operator.12 Janki13 in his study stated that the musculoskeletal complaints may lead to even work absence. Musculoskeletal disorders are a significant occupational hazards in conventional medical practice.14, 15 Block performance doesn’t represent an occupational hazard to most anaesthesiologists. However such hazards may apply to those with preexisting back or neck condition or to those who perform many more blocks.16 Our study differs from other studies where they use phantoms; the experience of operator who participated had varied experience. The experience of anaesthetists in our study is uniform, they had done more than 20 USG procedure and the study was conducted on patients who under went hernia repair under spinal anaesthesia. Advantage of our study is same experienced anaesthetists performed procedure in two groups. Limitation of our study is participants knew that they were observed during the procedure.

Conclusion

Safe practice of UGRA requires both knowledge and practical skills. In all our cases monitor position didn’t affect the success of the block, but keeping the monitor in front improves the efficacy and safety of the procedure. It improves ergonomic performance and reduces the procedure time significantly which is always preferred when many number of cases are being done.

Source of Funding

None.

Conflict of Interest

None.

References

1 

M S Abrahams M F Aziz R F Fu J L. Horn Ultrasound guidance compared with electrical neurostimulation for peripheral nerve block: a systematic review and meta-analysis of randomized controlled trialsBr J Anaesth20091023408170007-0912Elsevier BV

2 

Anjalee Brahmbhatt Michael J. Barrington Quality Assurance in Regional Anesthesia: Current Status and Future DirectionsCurr Anaesthsiol Rep201334215222167-6275Springer Science and Business Media LLC

3 

Michael J. Barrington Roman Kluger Ultrasound Guidance Reduces the Risk of Local Anesthetic Systemic Toxicity Following Peripheral Nerve BlockadeReg Anesth Pain Med2013384289971098-7339BMJ

4 

Joseph B Mabula Phillipo L Chalya Surgical management of inguinal hernias at Bugando Medical Centre in northwestern Tanzania: our experiences in a resource-limited settingBMC Res Notes2012515851756-0500Springer Science and Business Media LLC

5 

A O Oscar David G S Juan Camilo C Doris A Mauricio R M Angela Maria TAP block in inguinal hernia repairRandomized Controlled Trial Rev Colomb Anestesiol201745315965

6 

Ahmet Topal Mehmet Sargin Alper Kilicaslan Sema Tuncer Uzun. The effect of transverse abdominis plane block in inguinal hernioplasty on chronic painEur J Gen Med20151242917

7 

Getúlio Rodrigues de Oliveira Filho Pablo Escovedo Helayel Diogo Brüggemann da Conceição Ivo Sebastião Garzel Patrícia Pavei Maurício Sperotto Ceccon Learning Curves and Mathematical Models for Interventional Ultrasound Basic SkillsAnesth Analg20081062568730003-2999Ovid Technologies (Wolters Kluwer Health)

8 

G. A. Chapman D. Johnson A. R. Bodenham Visualisation of needle position using ultrasonographyAnaesth2006612148580003-2409, 1365-2044Wiley

9 

R A Langford B Hockey K Leslie Martuar position and accuracy and speed of ultrasound guided nerve backsAnaesth2009648459

10 

M Speer N Mclennan C Nixon Novice learner in plane ultrasound imaging; which visualization techniqueReg Anaesth Pain Med2013383502

11 

J M Wilson G Germain H Vaghadhia R Tang A Sawka In plane ultrasound guide needle insertion along or across the visual axis hand positionBr J Anaesth20141137188

12 

A K Hemal M Srinivas A R Charles Ergonomic Problems Associated with LaparoscopyJ Endourol20011554995030892-7790, 1557-900XMary Ann Liebert Inc

13 

Shiromani Janki Evalyn E. A. P. Mulder Jan N. M. IJzermans T. C. Khe Tran Ergonomics in the operating roomSurg Endosc20173162457660930-2794, 1432-2218Springer Science and Business Media LLC

14 

Derek R. Smith Ning Wei Lin Kang Rui-Sheng Wang Musculoskeletal disorders among professional nurses in mainland ChinaJ Prof Nurs20042063905

15 

M D Anghel D Lungeanu V Argesanu C A Comes C I Talpos Evidence of the need for ergonomic training of the dentists; there are cost of flawed posture and working routine (Abstract)Eur Cells Mater2008165

16 

Muhammad Ajmal Susan Power Tim Smith George D. Shorten Ergonomic task analysis of ultrasound-guided femoral nerve block: a pilot studyJ Clin Anesth201123135410952-8180Elsevier BV



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.